TESATRONIC TWIN-T10 Probe Display Unit - 04430013
Portable Display for Inductive Probe
Compact and heavy-duty instrument designed for the shop-floor.
TWIN-T10 portability provides great operational flexibility on the shop-floor. A robust internal battery provides up to 400 hours of continuos use, virtually eliminating concerns over unplanned downtime.
The carefully designed user interface supports a positive user experience and helps prevent operating errors. A large, high contrast display makes it easy to read measured results and wide input keys provide a crisp tactile response to the users touch.
The high-quality elastomer overmolding features a softtouch finish, making the instrument wear-resistant and comfortable to hold. Superior protection against dust and particles in production and assembly areas in ensured thanks to its IP63 rating.
The large high-contrast display and the shape of the analogue scale have been specifically designed to visualize easily slight variations in form or to detect turnaround point.
TWIN-T10 is particularly suitable for straightness, run-out, form or geometry evaluation during fine adjustment, alignment or fitting of mechanical parts.
Application Examples:
- Squareness adjustment of the Z-axis of a machine
- Alignement of guide rail to the marble bench
- Centering of a part on machine-tool
- Measurement of parallelism and perpendicularity
Features:
- Autonomous instrument used during assembly, on an inspection workstation of a production line, for final inspection or directly on a machine on the shop floor
- Frequently used with a GT 31 lever probe for geometry measurements: form tolerances (straightness, flatness etc.) or orientation tolerances (parallelism, perpendicularity, etc.)
- Function TOL for measurements with tolerances
- Memory function for values MAX, MIN or MAX-MIN for dynamic measurements
- Function for zero-setting of the display, for easy comparative measurements with a reference part
- Special ZOOM mode for a more detailed visualization of the analogue scale. This mode simplifies the alignment and fine adjustement during assembly.
- 4 or 7 measuring ranges from ± 5 µm to ± 5 mm, or switchable automatically depending on the measured value
- Access to functions by direct keys
- Millimetre/inch conversion
- 1 probe signal input
- Power supply by standard AA batteries
- RS232 digital output (TLC connector)
Technical Data:
Order number | 04430013 | |
Function | Display for TESA Inductive Probe | |
Number of probe entry | 1 | |
Display scales | ± 5, ± 2 mm ± 500, ± 200 μm ± 50, ± 20 μm ± 5 μm Auto Range | ± 250 ± 100 in/1000 ± 25 ± 10 in/1000 ± 2.5 ± 1.0 in/1000 ± 0.25 in/1000 Auto Range |
Resolution | 1 0,1 μm | 0.010 0.005 in/1000 |
Functions | Zeroing (Offset) +A, -A MAX, MIN or MAX-MIN memorization value Measurement with tolerances | |
Unit | Metric (mm, μm) or inch (in/1000) | |
Autonomy | 340 – 400 hours | |
IP rating | IP63 | |
Deviation span of indication * | ≤ 1% of the measuring scale | |
Drift of zero point on the full measuring range * | ≤ ± 0,005% / °C | |
Response time | ≤ 100 ms | |
Frequency limit | 10 Hz (relative to the input signal) | |
Oscillator frequency | 13 kHz | |
Digital output | RS232 through TLC connector | |
Data format | 7 bits ASCII code | |
Power supply | 4x AA LR6 batteries | |
Assigned operating T°C | + 20 °C ± 1 | |
Operating T°C range | +10 to +40 °C | |
Storage T°C range | -10 to +60 °C | |
EMC compatibility | according to 2004/108/EC according to EN 61326-1 annex A | |
Others: RoHS 2 REACH WEEE | according to 2011/65/EU according to EC 1907/2006 according to 2002/96/EC | |
Dimensions | 170 x 100 x 38 mm | |
Weight | 410 g / 500 g | |
Delivery contents | TWIN-T10, user manual, 4x AA batteries |
For more info, visit www.p65warnings.ca.gov.